Skip to product information
1 of 2
Nova Analytical Systems →

5460 SERIES PORTABLE ENGINE EXHAUST ANALYZERS

5460 SERIES PORTABLE ENGINE EXHAUST ANALYZERS

SKU:BYP-32312a

N/A

Regular price $49.99
Regular price Sale price $49.99
Sale

Initiate Order

5460 SERIES PORTABLE ENGINE EXHAUST ANALYZERS

5460 SERIES PORTABLE ENGINE EXHAUST ANALYZERS

$49.99

Quantity: 1

Claim Your Seller Account

5460 SERIES PORTABLE ENGINE EXHAUST ANALYZERS

5460 SERIES PORTABLE ENGINE EXHAUST ANALYZERS

$49.99

Quantity: 1

Delivery via Maden

Expect your order to arrive on time.

Secure Payments

All orders are processed through a secure, PCI-compliant checkout.

View full details
The Nova 5640 Series Portable Engine Exhaust Analyzers are built on the NOVANOW next-generation rugged portable platform. This series has been designed for accuracy, reliability, ease of use and ease of service, providing a detailed spot-check analysis of engine exhaust gas composition. The 5460 uses customer-replaceable electrochemical sensors that respond quickly to the O2, ppmCO, NO, NO2, and SO2 present in the exhaust gas sample. The O2 sensor life expectancy is typically 3-4 years. The CO, NO, NO2, and SO2 sensor life is typically 2-3 years. The infrared CO2 and HCs sensors should not need to be replaced under normal conditions of use.A rechargeable lithium ion battery provides enough power for about 8 hours of continuous operation and the analyzer can be used while it is being recharged. The 5460 series integrated case is small but rugged, and dust & water resistant when closed. The 5460 Series Analyzers are complete with built-in sampling pump, flow meter, automatic condensate removal, filters, PTFE liquid blocker, 12 ft. (4m) sampling hose, and stainless steel engine exhaust probe with pre-filter. Also included are a cigarette lighter power plug and remote printer actuator button for printer-equipped models.

Product Specifications

Applications 0
For spot-checking engine exhaust emissions for diesel, gasoline, propane, or natural gas powered 2 and 4 cycle engines. Analysis of carbon monoxide (CO), hydrocarbons (HC’s), carbon dioxide (CO2), oxygen (O2), nitric oxide (NO), nitrogen dioxide (NO2), and sulfur dioxide (SO2). For spot-checking engine exhaust emissions for diesel, gasoline, propane, or natural gas powered 2 and 4 cycle engines. Analysis of carbon monoxide (CO), hydrocarbons (HC’s), carbon dioxide (CO2), oxygen (O2), nitric oxide (NO), nitrogen dioxide (NO2), and sulfur dioxide (SO2).
Features 0
• Same rugged easy-to-use design as traditional Nova portable analyzers, but approx. 65% smaller and 50% lighter• Fast warm up and accurate response• Rugged touch-screen display of gas readings• Solid state infrared detectors for % CO, CO2, and HC's• Electrochemical sensors for O2, ppmCO, NO, NO2, and SO2• Rechargeable lithium ion battery; up to 8hrs use per charge (recharge from vehicle or from AC adapter)• Provided with sample filter, and flexible sample probe with 12 ft. of hose, built-in sample pump and automatic condensate removal• Local USB data logging creates CSV files• Built-in trouble-shooting prompts• Meets Bar 97 and ISO 3930/OIML R99, Class O specifications
Options 0
• 4-20mA output in external adaptor• Built-in printer, with remote print button• PPM CO in place of % CO measurements for recent engine designs with lower emissions
Specifications Accuracy Repeatability
±1% of full scale for O2, CO, CO2, and HCs; ±2% of full scale for NOx and SO2
Specifications Ambient Temperature Range
2°C to 40°C (35°F to 113°F) ºC
Specifications Analyzer Type
Portable Engine Exhaust Analyzer
Specifications Drift
< ± 1% in 8 hours after full warmup
Specifications Linearity
±1.0% of full scale each gas measured
Specifications Method Of Detection
CO (low range), NO, NO2, SO2, and O2 by electrochemical sensor
Specifications Options
built-in printer
Specifications Power
115 VAC or 220VAC, 50/60Hz. Also 12VDC for cigarette lighter socket
Specifications Primary Gases To Be Measured
carbon monoxide (CO), hydrocarbons (HC’s), carbon dioxide (CO2), oxygen (O2), and nitric oxide (NO).
Specifications Ranges Available
0-2,000 / 5,000 / 10,000 HCs (as hexane) | 0-2,000 / 5,000 / 10,000 PPM CO (low range) | 0-2,000 / 5,000 PPM NOx (as NO) | 0-2,000 PPM SO2 | 0-20,000 PPM HC's (as propane) | 0-20.0% CO2 | 0-25.0% O2 | 0-800 PPM NO2
Specifications Resolution
0.1% on % ranges, 0.01% on standard range CO, 1 PPM on most PPM ranges
Specifications Response Time T 90
usually 8-10 seconds
Specifications Size Weight
approx. 26 x 18 x 13 cm @ 3.4 kg (10" x 7" x 5" @ 7.5lbs)