Atlas Material Testing Technology LLC →

Atlas SUNTEST CPS+ Xenon Weathering Test Chamber

Atlas SUNTEST CPS+ Xenon Weathering Test Chamber

SKU:SUN-88990f

Enhance material testing with the Atlas SUNTEST CPS+ Xenon Chamber. Premium quality, precise results.

Regular price $9,999.00
Regular price Sale price $9,999.00
Sale

Initiate Order

Atlas SUNTEST CPS+ Xenon Weathering Test Chamber

Atlas SUNTEST CPS+ Xenon Weathering Test Chamber

$9,999.00

Quantity: 1

Claim Your Seller Account

Atlas SUNTEST CPS+ Xenon Weathering Test Chamber

Atlas SUNTEST CPS+ Xenon Weathering Test Chamber

$9,999.00

Quantity: 1

Delivery via Maden

Expect your order to arrive on time.

Secure Payments

All orders are processed through a secure, PCI-compliant checkout.

View full details
The Atlas SUNTEST CPS+ Xenon Weathering Test Chamber is an essential tool for testing polymers, cosmetics, packaging, and printing materials. With an exposure area of 560 square centimeters, equipped with powerful 1500W air-cooled Xenon lamps, this chamber offers premium simulation of solar radiation. It features microprocessor control, direct setting of irradiance and temperature, and a user-friendly interface with multilingual options. Meeting ASTM, ISO, and COLIPA standards, this chamber ensures accurate testing for UV-stability, color stability, and more, making it ideal for industrial applications requiring precise weathering and lightfastness testing.

Product Specifications

Accessories
SunFlood immersion unit for testing materials under extreme moisture conditionsSunCool chiller for testing at low temperatures (e.g. cosmetics)Daylight filterWindow Glass filterSolar ID65 filterStore Light filterSunCool Chiller for testing at low temperaturesSunFlood:40 liter temperature controlled immersion system for testing the weather resistance of materials, e.g. paints, plastics or building materialsSunTray:SunTray sample exchange unit including sample holder for 8 PMMA platesWater-cooled specimen table for exposure of thermo-sensitive samplesICH-D10/20/30 powder sample holders for preparation of 1-3 mm powder samples (Pharmaceuticals)SunCal BB 300-400 BST sensor for calibration of irradiance in W/m² in the wavelength range 300-400 nm and for calibration of BST in ºCSunCal WB 300-800 BST sensor for calibration of irradiance in W/m² in the wavelength range 300-800 nm and for calibration of BST in ºCSunCal Lux BST - Calibrated Lux-Sensor for calibration (via IBM compatible PC) of illuminance (lx) and radiant exposure (lxh), combined with calibrated BST sensor for calibration of Black Standard Temperature in ºC
Applications
Additives & ColorantsCosmeticsGraphic ArtsPackagingPharmaceuticalsPlastics
Benefits
Premium simulation of solar radiation according to CIE85 reference sunOptional simulation of window glass filtered daylight, and othersUser friendly interface with 11 selectable languagesPre-programmed 300 hour tests for Quick-start setup or as templates for custom testsBlack Standard Temperature Sensor (BST) according to ISO 4892-1Automatic control of irradiance for various testing needs
Features
Microprocessor control, with programming keypad and black/white displayDirect setting and control of irradiance at 300-400 nm or 300-800 nmDirect setting and control of Black Standard Temperature (BST)Display of chamber air temperature (CHT)Display of test values and diagnostic messagesParameter checkTwo pre-programmed test methods plus storage for six user-defined testsMultilingual: English, German, Italian, French, Spanish, Polish, Russian, ChineseHorizontal test area max. 560 cm² for flat samples or 3D specimensPremium sunlight simulation according to CIE85 reference sunNon-aging coated quartz filter system and selectively reflecting mirrorsRead to accommodate additional optical filtersSwitch-off modes: radiant exposure or test timeData output via serial interface RS232
Options
Light Monitor 300-400 nmLight Monitor 300-800 nmHumidity Sensor for measuring relative humidity during testing
Overview
The practical benchtop xenon lamp unit with useful accessories to tailor your SUNTEST towards your needs. Here you see the SunCool chiller (for pharmaceuticals / cosmetics), the SunTray sample exchanger (for in-vitro SPF), and the SunFlood immersion unit ( for coatings / sealants).The SUNTEST CPS+ is used to test for property changes of materials caused by sunlight, temperature and moisture in a short period of time. Product changes happening outdoor or indoor over months or years, such as fading, yellowing or embrittlement can often be simulated within weeks inside a SUNTEST.
Specifications Black Standard Temperature Bst Range Without Sunc
35 to 100 ºC
Specifications Depth
35 cm
Specifications Exposure Area
28 x 20 cm | 560 cm²
Specifications Height
35 cm
Specifications Irradiance Range
250 to 765 W/m² | 30 to 65 W/m² | 300 to 400 nm | 300 to 800 nm | 340 nm
Specifications Lamp Phase
1
Specifications Lamp Power
1500 W
Specifications Light Source
Air-Cooled Xenon Lamps
Specifications Software
XenoSoft
Specifications Width
90 cm
Standards
The listed standards and specifications may include several alternative testing options. The instrument meets at least one option in any listed standard. Standards are subject to change without notification, which may result in a change in testing technology. Always check the revision status of a standard with the responsible Standards Development Organization. If unsure about instrument compliance, please contact your local sales representative.ASTM C1442Standard Practice for Conducting Tests on Sealants Using Artificial Weathering ApparatusASTM C1501Standard Test Method for Color Stability of Building Construction Sealants as Determined by Laboratory Accelerated Weathering ProceduresASTM D3424Standard Practice for Evaluating the Relative Lightfastness and Weatherability of Printed MatterASTM D5071Standard Practice for Exposure of Photodegradable Plastics in a Xenon Arc ApparatusASTM D6695Standard Practice for Xenon-Arc Exposures of Paint and Related CoatingsASTM G151Standard Practice for Exposing Nonmetallic Materials in Accelerated Test Devices that Use Laboratory Light SourcesASTM G155Standard Practice for Operating Xenon Arc Light Apparatus for Exposure of Non-Metallic MaterialsCOLIPA in-vitro UVA (2011)Method for in vitro determination of UVA protectionEPA/ASTM E896Standard Test Method for Conducting Aqueous Direct Photolysis TestsGB/T 8427Textiles - Tests for color fastness - Color fastness to artificial light: Xenon arc fading lamp testGB/T 16422Plastics - Methods of exposure to laboratory light sources -Part 2: Xenon-arc lampsICH Q1BQ1B Photostability Testing of New Drug Substances and ProductsICH Q5CStability testing of Biotechnological / Biological productsISO 4049Dentistry - Polymer-based restorative materialsISO 7491Dental materials - Determination of Color stabilityISO 4892-1Plastics - Methods of exposure to laboratory light sources - Part 1: General guidanceISO 4892-2Plastics - Methods of exposure to laboratory light sources - Part 2: Xenon-arc sourcesISO 10977Photography - Processed photographic Color films and paper prints - Methods for measuring image stabilityISO 11431Building construction - Jointing products - Determination of adhesion/cohesion properties of sealants after exposure to heat, water and artificial light through glassISO 11979-5Ophthalmic implants - Intraocular lenses - Part 5: BiocompatibilityISO 24443Determination of sunscreen UVA photoprotection in vitroL’OREAL QAC-MC-151-OAccelerated ageing under lightOECD 495Reactive Oxygen Species (ROS) Assay for PhotoreactivityOECD 432OECD Guideline for testing of chemicals. In Vitro 3T3 NRU phototoxicity testVICH GL5Guideline On Stability Testing: Photostability Testing of New Veterinary Drug Substances and Medicinal Products