High-Speed Transient Recording I/O Modules
High-Speed Transient Recording I/O Modules
SKU:TRA-837eb2
Capture fast events at up to 10 MS/s with our industry-leading Transient Recording I/O Modules. Unparalleled data acquisition efficiency.
Regular price
$3,299.99
Regular price
Sale price
$3,299.99
Unit price
per
Delivery via Maden
Expect your order to arrive on time.
Secure Payments
All orders are processed through a secure, PCI-compliant checkout.
Experience unparalleled data acquisition with our High-Speed Transient Recording I/O Modules. Capable of recording fast events at up to 10 MS/s, these modules feature rugged enclosures accommodating 3 or 16 channels. With dual inputs including bridge and ICP compatibility, programmable for various bridge types, and automatic hands-off operation, these modules ensure precise and efficient signal processing. The Ethernet interface, along with Windows software, allows seamless data transfer and management. Ideal for applications requiring high-speed event capture and long-term data storage.

Product Specifications
Amplifier Bandwidth
5841/43 - 100 kHz. Slew rate is 5 V/µS.5842/44/45/46 - 500 kHz for gains below 100, 300 kHz for higher gains. Slew rate is 15 V/µS.
Amplifier Cm Voltage
±10 V
Amplifier Common Mode
60 dB plus gain in dB to 110 dB minimum for 350 Ohm source unbalance, DC to 60 Hz.
Amplifier Gain
Programmable steps 1, 2, 3, 5, 10, 20, 30, 50, 100, 200, 300, 500 and 1,000.
Amplifier Gain Linearity
±0.02 %
Amplifier Gain Stability
±0.02% for 8 hours, ±0.005%/ºC.
Amplifier Input Impedance
25 Megohms shunted by 500 pF.
Amplifier Input Protection
±50 Volts differential or common mode.
Amplifier Input Range
±10 mV to ±10 Volts full scale.
Amplifier Noise
10 µV RTI plus 1 mV RTO RMS.
Amplifier Zero Stability
±5 µV RTI, ±1 mV RTO at constant temperature. ±1 µV/ºC RTI, ±0.2 mV/ºC RTO.
Bridge Input Balance Accuracy
±0.1% of range, ±1 mV
Bridge Input Balance Range
2 mV/V for 350 Ohm bridge and is setable by resistor from < 0.1 mV/V to > 10 mV/V.
Bridge Input Balance Resolution
0.025 %
Bridge Input Bridge Balance
Automatic, may be turned off without affecting the setting.
Bridge Input Bridge Completion
Programmable completion is installed for 350 Ohm and 120 Ohm 1/4 and all 1/2 bridge circuits. The 1/4 bridge completion resistors are in resolderable sockets and may be replaced to accommodate other values.
Bridge Input Bridge Connector
Miniature, 8-pin, threaded, round DIN. 5871ST has six-position terminal strip.
Calibration Shunt Calibration
Four-step shunt calibration. Resistor values of 150K, 75K, 15K, and 7.5K ohms give 0.583, 1.16, 5.77, and 11.4 mV/V for a 350 Ohm bridge, ±1%. Resistors are mounted in resolderable terminals. Alternate values may be specified.
Calibration Voltage Calibration
Three ranges, ±10 Volts, ±1 Volt and ±100 mV. 0.025% of range (12-bits) of range with ±0.25% Accuracy.
Digitizer Correlation
1 nS sample to sample, 50 nS channel to channel.
Digitizer Linearity
5841 - 3.5 LSB5842, 5844, 5845 & 5846 - 2.5 LSB
Digitizer Rates
5841: 1M, 500K, 200K, 100K, 50K, 20K, 10K, 5K, 2K, 1K, 500, 200, 100, 50, 20 and 10 S/s.5842: 2M, 1M, 500K, 200K, 100K, 50K, 20K, 10K, 5K, 2K, 1K, 500, 200, 100, 50, and 20 S/s.5843: 200K, 100K, 50K, 20K, 10K, 5K, 2K and 1K S/s.5844/45/46: 10M, 5M, 2M, 1M, 500K, 200K, 100K, 50K, 20K, 10K, 5K, 2K, 1K, 500, 200 and 100 S/s.
Digitizer Resolution
5843 - 24-bits5841 & 5846 - 16-bits5842, 5844 & 5845 - 14-bits
Digitizer Sampling
One pre-trigger and two post-trigger sample rates per record.
Filter Filter Type
5841/42/44/45/46 - Eight-pole, Bessel, low-pass (48 dB/octave terminal slope). | 5841/43 - Standard is 5 kHz, 10 kHz, 20 kHz, 50 kHz (-3 dB) and wideband. | 5842/44/45/46 - Standard is 10 kHz, 20 kHz, 50 kHz, 100 kHz (-3 dB) and wideband. Other filter frequencies are available on request.
Iepe Input Ac Coupling
Programmable high pass filter, 2.25 Hz. Input impedance is 100K Ohms.
Iepe Input Iepe Connector
BNC
Iepe Input Iepe Excitation
6 mA, ±10% (range is 1 to 20 mA).
Iepe Input Iepe Filter
Selectable low pass input filter, 16 kHz.
Iepe Input Iepe Input
AC coupled input (< 2.25 Hz) to accept charge transducer with integral or line preamplifier.
Iepe Input Iepe Input Impedance
100K Ohms
Input Configuration
Dual Input, Bridge and ICP/IEPE with analog outputs
Input Input Stability
±0.1% for 8 hours, ±0.01%/ºC
Memory Records
Programmable pre and post-trigger data in 4096 word blocks. Multiple records without debriefing.
Memory Size
3.5M words (24-bit) to 8M words (14 or 16-bit)
Memory Type
CMOS, with power back up for 40 hours.
Voltage Excitation Transducer Power Voltage Excitation
0 to 12 Volts in 1-Volt ±0.1% steps. 50 mA, limited to 70 mA maximum.
Voltage Excitation Transducer Power Voltage Excitation Regulatio
±0.2% over the input voltage range and no load to full load.
Voltage Excitation Transducer Power Voltage Monitor
Excitation voltage and current is digitized output and recorded. Accuracy is ±0.1%.
Voltage Excitation Transducer Power Voltage Stability
±0.1% for 8 hours, ±0.01%/ºC.