The Nova 207RT Series Portable Dew Point Transmitter have been designed primarily for accurate dew point measurement of atmosphere gas flowing from endothermic gas generators, at slightly positive pressure.The 207RT may be mounted anywhere on the gas main between the generator and the furnace. The dielectric polymer sensor provides stable and accurate dew point measurements and is not affected by most gas compositions. (some exceptions)Because there is no built-in sample pump to draw in a sample, the positive pressure produced by the generator pushes the sample into the flow-through chamber. The sample flow is shown on a small rotameter with valve. The analyzer then reads the dew point continuously.
Product Specifications
Applications 0
For continuous dew point monitoring of atmospheric gas from endothermic generators.
Calibration 0
Calibration is easily performed in the field using the optional calibration kit (Part no. NDPCK-02) Alternatively, the analyzer may be returned to Nova for a certified calibration
Features 0
Standard range of -40 to +104°F (-40 to +40°C) Dew Point No moving parts 4-20mA output to controllers and recorders Long-life dielectric polymer sensor May be calibrated in the field Bright digital readout Runs on 12 to 24VDC Mounts anywhere on gas main between generator and furnace Sensor easily replaced if contaminated Factory calibrated to NIST traceable standards
Options 0
Display also available loose for remote panel mounting (207RT-PM) Available with display and sensor only - as spare during calibration or repair periods; no flow through chamber Calibration kit Power supply optionally available
Specifications Accuracy Repeatability
±1°F (±1°C) Dew Point
Specifications Alarms
High or low dew point alarms with LED (optional)
Specifications Ambient Temperature Range
-18 to 49 ºC | 0 to 120 ºF
Specifications Drift
Less than 1% of full scale range per month
Specifications Hazardous Area Rating Options
General Purpose, Non-Hazardous
Specifications Linearity
Better than 1°F (1°C) Dew Point
Specifications Method Of Detection
Dielectric polymer sensor that change capacitance with moisture variation
Specifications Output Options
4-20 mA
Specifications Power
12 to 24 VDC
Specifications Ranges Available
-40°C to +40°C Dew Point | -40°F to +104°F Dew Point
Specifications Resolution
1°F (1°C) Dew Point
Specifications Response Time T 90
Less than 30 seconds
Specifications Size Weight
Approx. 10" L x 6" W x 4" D @ 5 lbs. (26 x 25 x 10 cm @ 2.3 kg)
Unique Applications 0
The 207RT Series Dew Point Transmitter should not be used to measure dew points when trace amounts of sulfur trioxide (SO3), chlorine (Cl2), hydrogen chloride (HCl), or ammonia (NH3) are present. All Nova analyzers are built using proven technologies and techniques. If this product does not suit your application, please contact Nova. In many cases, we are able to build an analyzer specific to your needs.