Skip to product information
1 of 2
Nova Analytical Systems →

Model 410 D1 or D2 Oxygen Analyzer with cabinet purge systems for hazardous areas

Model 410 D1 or D2 Oxygen Analyzer with cabinet purge systems for hazardous areas

SKU:BYP-df1263

N/A

Regular price $49.99
Regular price Sale price $49.99
Sale

Initiate Order

Model 410 D1 or D2 Oxygen Analyzer with cabinet purge systems for hazardous areas

Model 410 D1 or D2 Oxygen Analyzer with cabinet purge systems for hazardous areas

$49.99

Quantity: 1

Claim Your Seller Account

Model 410 D1 or D2 Oxygen Analyzer with cabinet purge systems for hazardous areas

Model 410 D1 or D2 Oxygen Analyzer with cabinet purge systems for hazardous areas

$49.99

Quantity: 1

Delivery via Maden

Expect your order to arrive on time.

Secure Payments

All orders are processed through a secure, PCI-compliant checkout.

View full details
The Nova 410 Series represent Nova's best value in continuous process O2 analysis. This analyzer utilizes a long-life electrochemical 'fuel cell' sensor for the detection of Oxygen in process gases. The sensor produces a small millivoltage output which is directly proportional to the Oxygen detected. The output is then amplified and displayed on a digital readout meter and can also be sent to an optional recorder output. The 410 can be supplied with sampling pump or pressure regulator to suit most applications.Cabinet: NEMA 4 cabinet with Hazardous Area Purge System; available for Class 1 Div1 & Div2 hazardous areas.

Product Specifications

Applications 0
For continuous analysis of oxygen (O2) in process gas streams. Also recommended for other applications such as O2 enrichment of combustion air, flue gas analysis, oxygen deficiency analysis, and many more.
Calibration 0
Nitrogen for zero Air or analyzed calibration gas for span
Features 0
Best value continuous O2 analyzer Bright digital readout Long life electrochemical O2 sensor Fast response (T90 - 6 to 8 seconds) Microprocessor control for push button calibration Easy-to-maintain modular layout Built in sample pump or pressure regulator 4-20mA outputs
Options 0
Hi/Low gas, low flow, and diagnostic alarms available Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available Light-duty sample conditioning available for removal of moisture and dust Cabinet purge system available for use in hazardous areas Cabinet cooling for areas not air-conditioned Sensors mounted in temperature controlled heated box to improve accuracy High temperature filters and probes
Specifications Accuracy Repeatability
Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale
Specifications Alarms
High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating | Low flow alarm
Specifications Ambient Temperature Range
32 to 122°F (0 to 50°C), Lower temperatures (-5°F, -20°C) with Cold Weather Package
Specifications Analyzer Type
O2 Analyzer
Specifications Drift
Less than 2% of full scale per month
Specifications Hazardous Area Rating Options
Class 1 Div 2 or Class 1 Div 1 Hazardous Area Ratings available
Specifications Linearity
Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale
Specifications Method Of Detection
Customer replaceable electrochemical 'fuel cell' O2 sensor. Expected life 3 to 5 years depending on typical measurement concentrations.
Specifications Output Options
4-20 mA into 500 ohms non-isolated standard | Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
Specifications Power
115 VAC, 60 Hz | 220 VAC, 50 Hz
Specifications Primary Gases To Be Measured
Oxygen (O2)
Specifications Ranges Available
Any range from 0-2.0% to 0-100.0% O2
Specifications Resolution
0.1% O2
Specifications Response Time T 90
6-8 seconds to 90% step change
Specifications Size Weight
Dimensions will vary depending on type of cabinet purge required | NEMA 4: 51H x 41W x 26D cm (20"H x 16"W x 10"D)