Skip to product information
1 of 2
Nova Analytical Systems →

Model 410N7MC Process Oxygen Analyzer, in Explosion-proof cabinet, Class 1 Div 1 Group BCD

Model 410N7MC Process Oxygen Analyzer, in Explosion-proof cabinet, Class 1 Div 1 Group BCD

SKU:BYP-6dfa2c

N/A

Regular price $49.99
Regular price Sale price $49.99
Sale

Initiate Order

Model 410N7MC Process Oxygen Analyzer, in Explosion-proof cabinet, Class 1 Div 1 Group BCD

Model 410N7MC Process Oxygen Analyzer, in Explosion-proof cabinet, Class 1 Div 1 Group BCD

$49.99

Quantity: 1

Claim Your Seller Account

Model 410N7MC Process Oxygen Analyzer, in Explosion-proof cabinet, Class 1 Div 1 Group BCD

Model 410N7MC Process Oxygen Analyzer, in Explosion-proof cabinet, Class 1 Div 1 Group BCD

$49.99

Quantity: 1

Delivery via Maden

Expect your order to arrive on time.

Secure Payments

All orders are processed through a secure, PCI-compliant checkout.

View full details
The Nova 410 Series represent Nova's best value in continuous process Oxygen analysis. This analyzer utilizes a long-life electrochemical 'fuel cell' sensor for the detection of O2 in process gases. The sensor produces a small millivoltage output which is directly proportional to the Oxygen detected. The output is then amplified and displayed on a digital readout meter and can also be sent to an optional recorder output. The 410 can be supplied with sampling pump or pressure regulator to suit most applications.Cabinet: Wall Mounted NEMA7 UL/CSA Explosion-proof with Non-intrusive Magnetic Calibration, Class 1 Div 1 Group BCD. Availability depends on application and environment. Sample conditioning options not available with this cabinet style.

Product Specifications

Applications 0
For continuous analysis of oxygen (O2) in process gas streams. Also recommended for other applications such as O2 enrichment of combustion air, flue gas analysis, oxygen deficiency analysis, and many more.
Calibration 0
Nitrogen for zero Air or analyzed calibration gas for span
Features 0
Best value continuous O2 analyzer Bright digital readout Long-life electrochemical O2 sensor Fast response (T90 - 6 to 8 seconds) Touch-screen or magnetic wand calibration control, depending on cabinet configuration Easy-to-maintain modular layout Built-in sample pump or pressure regulator 4-20mA output
Options 0
Hi/Low gas, low flow alarms available Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available Sensors mounted in temperature controlled heated box to improve accuracy
Specifications Accuracy Repeatability
Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale
Specifications Alarms
High and/or Low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating | Low flow alarm
Specifications Ambient Temperature Range
32 to 122°F (0 to 50°C)
Specifications Analyzer Type
O2 Analyzer
Specifications Drift
Less than 2% of full scale per month
Specifications Hazardous Area Rating Options
Class 1, Div 1, Grp A-D Hazardous Area Rating
Specifications Linearity
Below 30% O2 ±1% of full scale; above 30% O2 ±2% of full scale
Specifications Method Of Detection
Customer replaceable electrochemical O2 sensor. Expected life 3 to 5 years depending on typical measurement concentrations.
Specifications Output Options
4-20 mA into 500 ohms non-isolated standard | Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
Specifications Power
115 VAC, 60 Hz | 220 VAC, 50 Hz
Specifications Primary Gases To Be Measured
Oxygen (O2)
Specifications Ranges Available
Any range from 0-2.0% to 0-100.0% O2
Specifications Resolution
0.1% O2
Specifications Response Time T 90
6-8 seconds to 90% step change
Specifications Size Weight
EX-PROOF: 61H x 33W x 31H cm (24"H x 13"W x 12"D)