The Nova 313 Series Portable Flue Gas Analyzer for NO, NO2, or NOx utilizes reliable, stable NO and/or NO2 sensors which respond quickly to the NO or NO2 present in flue gases or engine exhaust. NO is the major component (90-95%) of the NOx found in flue gases or engine exhaust (except diesel).In operation, a built in sample pump draws in the gas sample through the S.S. probe, 12 ft sample hose, condensate removal filter, secondary filter and flowmeter, then on to both sensors. The output of each sensor is then amplified and displayed on a digital meter.The rechargeable battery provides enough power for about 8 hours of operation and the analyzer can be used while it is being recharged. The recharger plus a stainless steel probe with 12 ft. (4 m) hose are included.
Product Specifications
Applications 0
Analysis of oxides of nitrogen (NOx) such as nitric oxide (NO) and/or nitrogen dioxide (NO 2 ). For boiler, furnace, or engine exhaust monitoring and analysis.
Calibration 0
Air for zero. Analyzed calibration gas mixtures of PPM NO and PPM NO2 in nitrogen for span.
Features 0
NOx can be read as NO, NO2, or as total Rugged design that is easy to operate Disposable, long life electrochemical NO and NO2 sensors Digital meter readout with backlight Modular layout that is easy to maintain Rechargeable battery operation Built in sample pump, filter, and flowmeter Active condensate removal Rapid reading recovery after NOx 'overdose' Suitcase cabinet (K) weather-proof when closed Stainless steel probe with sample hose
Options 0
Recorder output - 4-20mA Stack temperature readout (313T) Special sampling probes Sample pre-cooler Suitcase (K) style cabinet available On-board data-logging to USB stick
Specifications Accuracy Repeatability
Better than 2% full scale
Specifications Alarms
NO, NO2 or NOx alarms with LED
Specifications Ambient Temperature Range
12 to 49 ºC | 55 to 120 ºF
Specifications Drift
Within 1% of full scale per 8 hours of continuous operation
Specifications Hazardous Area Rating Options
General Purpose/Non-hazardous
Specifications Linearity
±2% of full scale
Specifications Method Of Detection
Customer replaceable electrochemical nitric oxide and nitrogen dioxide sensors Expected life 2-3 year each
Specifications Output Options
4-20mA
Specifications Power
12V battery operation. 115VAC 60Hz for recharging (220VAC 50Hz available)
Specifications Ranges Available
0-2000 PPM NO; 0-800 PPM NO2; 0-2000 PPM Nox; (0-5000 PPM available) Ranges switch selectable
Specifications Resolution
1 PPM on 0-2000, 10 PPM on 0-5000
Specifications Response Time T 90
20-30 seconds
Specifications Size Weight
approx. 26 x 18 x 13 cm @ 3.4 kg (10" x 7" x 5" @ 7.5lbs)